Dynalab 4000-0000 Manuel d'utilisation
Page 2

[ 2 ]
BEFORE USE
If the equipment is not used in the manner described
in this manual the protection provided by the
equipment may be impaired.
N.B.
The Merit Water Still is classified as
“Permanently Connected Equipment” and should be
connected to the electricity supply by a qualified
electrician in the manner described in the electrical
installation section of this manual.
The Merit Water Still is designed to operate under the
following conditions:
❖ For indoor use only
❖ Ambient temperature range +5°C to +40°C
❖ Altitude up to 2000m
❖ Relative humidity not exceeding 80%
❖ Mains supply voltage fluctuations not greater than
±10% of nominal
❖ Overvoltage category II IEC664
❖ Pollution degree 2 IEC664
Limitations of use
This product meets the applicable E.C. harmonised
standards for radio frequency interference and may be
expected not to interfere with, or be affected by other
equipment with similar qualifications. We cannot be sure
that other equipment used in its vicinity will meet these
standards and we cannot guarantee that interference will
not occur in practice. Where there is a possibility that injury,
damage or loss might occur if equipment malfunctions due
to radio frequency interference, or for general advice before
use, please contact the Technical Service Department of
Bibby Sterilin Ltd.
Merit Water Still W4000 & W4000/EURO
Figure 1 General assembly
Condenser
Boiler
Thermostat reset buttons
Heater
Vent
Distillate
outlet
Straps
Stopcock
To drain
Constant level
control
Cooling water
inlet
Gaskets and
flange
The Merit Water Still complies with the
following European Directives:
89/336/EEC - E.M.C. DIRECTIVE
Amended by 92/31/EEC and 93/68/EEC
73/23/EEC - L.V.D. Amended by 93/68/EEC