English – Velleman PROBE60S Manuel d'utilisation
Page 2

ENGLISH
Oscilloscope Probe Kit
Introduction
The PROBE60S is a passive high impedance oscilloscope probe designed and calibrated for use
on instruments having an input impedance of 1 Mohm shunted by 25pF. However, it may be
compensated for use with instruments having an input capacitance of 10 - 35 pF. The probe
incorporates a two position slide switch in the head which selects attenuation of X1, X10 position.
Maintenance
Before dismantling any part of the probe, make sure it is disconnected from any voltage source.
The probe head can be detached from the cable assembly by unplugging the push fit BNC
connector on the probe head. This permits replacement of the cable or head assembly should
either part become damaged. The measuring tip is also replaceable. To replace a broken tip, hold
the black insulating part of the tip with pliers and pull it away from the probe head. Replace with a
new tip taking care to align with the inner contact.
Compensation Adjustment
In order to obtain the correct division ratio with each
oscilloscope, the attenuation network needs to be
adjusted. To compensate the probe to your
oscilloscope, apply a 1kHz square wave to the
probe tip, or connect to the cal socket on the
oscilloscope to display a few cycles of the
waveform.
And adjust the trimmer located in the BNC connector for a flat-topped square wave.
Specifications
Position x1
Attenuation Ratio
1:1
Bandwidth
DC tot 15 MHz
Rise
Time 27
ns
Input Resistance
1Mohm (oscilloscope input)
Input Capacitance
46 pF plus oscilloscope capacitance
Working Voltage
600 V DC incl. Peak AC derating with frequency
Position x10
Attenuation Ratio
1:10
Bandwidth
DC tot 60 MHz
Rise
Time 5.5
ns
Input Resistance
10 Mohm when used with oscilloscopes with 1Mohm input
Input Capacitance
Approx. 18pF
Compensation Range
10 to 35 pF
Working voltage
600 V DC incl. peak-AC derating with frequency (see Fig. 1)
Cable length
±140 cm