Secondary_value_1
Calibration
Lowest Point
Calibration
Highest Point
Sensor_value
cal_level_lo
cal_level_hi
Calibration
Lowest Point
Calibration
Highest Point
Sensor reference plane
cal_level_lo [%]
cal_level_hi [%]
Fig. 27: Réglage VEGASON 66
Liste de paramètre pour Device revision 3.0
La liste suivante vous indique les paramètres les plus importants et leur signification :
l
primary_value
-
This is the process value after adjustment and Linearization with the status of the
transducer block
l
primary_value_unit
-
Unit code of 'Primary_value'
l
secondary_value_1
-
Value after min./max.-adjustment (level + level offset). Selected as input to AIFB by
setting 'Channel' = 2. Unit derives from 'Secondary_value_1_unit'
l
secondary_value_1_unit
-
Unit code of 'Secondary_value_1'
l
secondary_value_2
-
Sensor value + sensor offset. Selected as input to AIFB by setting 'Channel' = 3. Unit
derives from 'Secondary_value_2_unit'
l
secondary_value_2_unit
-
Unit code of 'Secondary_value_2'
Annexe
VEGASON 66 - Fieldbus Foundation